• Title of article

    Improved electrochemical immunosensor for myeloperoxidase in human serum based on nanogold/cerium dioxide-BMIMPF6/l-Cysteine composite film

  • Author/Authors

    Lu، نويسنده , , Lingsong and Liu، نويسنده , , Bei and Li، نويسنده , , Shenfeng and Zhang، نويسنده , , Wei and Xie، نويسنده , , Guoming، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    6
  • From page
    339
  • To page
    344
  • Abstract
    An electrochemical immunosensing assay for myeloperoxidase (MPO) determination in human serum has been developed. Firstly, l-Cysteine was initially electropolymerized on an Au electrode to form l-Cysteine film. After that cerium dioxide (CeO2) dispersed in 1-butyl-3-methylimidazolium hexafluorophosphate (BMIMPF6) were immobilized on the l-Cysteine film. Then the negatively charged nanogold particles were adsorbed onto the membrane via the positive charge of CeO2, which aimed at assembling more antibody of MPO (anti-MPO). The resulting immunosensor showed a high sensitivity, broad linear response to the MPO concentration comprised between 10 ng/mL and 400 ng/mL with a detection limit of 0.06 ng/mL. Moreover, the surface morphology of the electrode was studied by means of a scanning electron microscope and the electrochemical properties of the fabricated immunosensor were further characterized by cyclic voltammetry. Also, factors influencing the performance of the resulting immunosensors were studied in detail.
  • Keywords
    Myeloperoxidase , Ionic liquids , L-cysteine , CeO2 , Immunosensor
  • Journal title
    Colloids and Surfaces B Biointerfaces
  • Serial Year
    2011
  • Journal title
    Colloids and Surfaces B Biointerfaces
  • Record number