Title of article :
The endothelialization and hemocompatibility of the functional multilayer on titanium surface constructed with type IV collagen and heparin
Author/Authors :
Zhang، نويسنده , , Kun and Li، نويسنده , , Jing-Mei and Deng، نويسنده , , Kun and Liu، نويسنده , , Tao and Chen، نويسنده , , Junying and Huang، نويسنده , , Nan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
10
From page :
295
To page :
304
Abstract :
The type IV collagen/heparin (IVCol/Hep) multilayer was developed on amino-silanized titanium (Ti) surface layer by layer self-assembly. Ti, TiOH, TiOHA and TiOHA(HC)3H were characterized by Fourier transform infrared spectroscopy (FTIR), water contact angle measurements and scanning electron microscopy (SEM), respectively. Alcian Blue 8GX staining and immunofluorescence staining were used to characterize the heparin (Hep) and type IV collagen (IVCol), respectively. The blood compatibilities of Ti and the treated Ti were evaluated by platelet adhesion test and clotting time using PRP. Blood compatibility tests reveal that the assembled functional multilayer displayed less platelets adhesion and prolonged APTTs time compared with the controlled Ti. Endothelial cells (ECs) culture results showed more attached and proliferated ECs on the TiOHA(HC)3H than that on Ti, especially compared with that on TiOH and TiOHA. Thus, the assembled Hep and IVCol multilayer can improve the cell compatibility and the blood compatibility. We anticipate that this IVCol/Hep functional multilayer will be beneficial to enhance the biocompatibility of the Ti-based biomaterial devices.
Keywords :
Functional multilayer , Type IV collagen , HEPARIN , Endothelialization , Blood compatibility , Mechanism
Journal title :
Colloids and Surfaces B Biointerfaces
Serial Year :
2013
Journal title :
Colloids and Surfaces B Biointerfaces
Record number :
1976684
Link To Document :
بازگشت