• Title of article

    Nafion covered lead film electrode for the voltammetric determination of caffeine in beverage samples and pharmaceutical formulations

  • Author/Authors

    Tyszczuk-Rotko، نويسنده , , Katarzyna and B?czkowska، نويسنده , , Ilona، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2015
  • Pages
    6
  • From page
    24
  • To page
    29
  • Abstract
    This paper presents a sensitive, selective and low-cost voltammetric method for the determination of caffeine using a Nafion covered lead film electrode. The sensor was prepared on a glassy carbon electrode modified with lead film recovered by a Nafion layer. Caffeine was accumulated and then oxidised at the modified electrode surface to produce two anodic peaks at 0.86 and 1.40 V (vs. Ag/AgCl) in 0.1 mol L−1 H2SO4 medium. The obtained detection limits for caffeine following 120 s of accumulation time were equal to 1.7 × 10−8 mol L−1 (for peak 1) and 2.2 × 10−7 mol L−1 (for peak 2). The method was successfully applied to determination of caffeine in tea, coffee, soft and energy drink samples as well as pharmaceutical formulation and the contents closely corresponded to those quoted by the manufacturer and those obtained by the reported spectrophotometric method.
  • Keywords
    Nafion film , Lead film electrode , Adsorptive stripping voltammetry , caffeine
  • Journal title
    Food Chemistry
  • Serial Year
    2015
  • Journal title
    Food Chemistry
  • Record number

    1979692