Title of article
A novel solid phase extraction – Ultra high performance liquid chromatography–tandem mass spectrometry method for the quantification of ochratoxin A in red wines
Author/Authors
Mariٌo-Repizo، نويسنده , , Leonardo and Kero، نويسنده , , Frank and Vandell، نويسنده , , Victor and Senior، نويسنده , , Adam and Isabel Sanz-Ferramola، نويسنده , , M. Bramanti and M. C. Cerutti، نويسنده , , Soledad and Raba، نويسنده , , Julio، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2015
Pages
6
From page
663
To page
668
Abstract
A novel and advanced technology on solid phase extraction column prior to liquid chromatography coupled to tandem mass spectrometry has been used for the determination of ochratoxin A in red wine samples. Due to the need of a reliable and rugged method according to current regulations and with the aim of minimize heuristic efforts associated with analytical method development, the statistical design of experiment was employed. On other hand, the method validation according to European Commission 2002/657/EC was achieved. The values obtained for decision limit (CCα), detection capability (CCβ), limits of detection and quantification were 0.07 μg L−1, 0.14 μg L−1, 0.13 μg L−1 and 0.41 μg L−1, respectively. The recoveries values were ranged from 95.7% to 107.2%. These values were compatible with the 2.0 μg L−1 maximum allowable concentration limit established by different international regulations.
Keywords
Ochratoxin A , Red wine , UPLC–MS/MS , Solid phase extraction
Journal title
Food Chemistry
Serial Year
2015
Journal title
Food Chemistry
Record number
1979954
Link To Document