Title of article :
A study on the selection of chemiluminescence system for the flow injection determination of the total polyphenol index of plant-derived foods
Author/Authors :
Nalewajko-Sieliwoniuk، نويسنده , , Edyta and Malejko، نويسنده , , Julita and ?wi?czkowska، نويسنده , , Marta and Kowalewska، نويسنده , , Agata، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2015
Abstract :
Different chemiluminescence systems based on luminol, permanganate, manganese(IV) and cerium(IV) reagents were compared regarding their sensitivity and selectivity to determine plant polyphenols. Among the seventeen systems tested, Mn(IV)-formaldehyde-hexametaphosphate was considered to be the most suitable for polyphenols detection. The developed flow injection method (FI-CL) based on enhancing effect of polyphenols on Mn(IV) chemiluminescence is characterised by low detection limit of gallic acid (0.02 μg L−1) and high precision (RSD = 1.7%). The calibration graph was linear from 0.1 to 100 μg L−1. The selectivity studies revealed that the FI-CL method ensures accurate determination of the total polyphenols content in food samples. The method was successfully applied to analysis of a variety of plant-derived foods (wine, tea, cereal coffee, fruit and vegetable juices, herbs and spices). The proposed method is superior to conventional spectrophotometric assays due to its higher sample throughput (195 samples h−1), simplicity, sensitivity and, above all, higher selectivity.
Keywords :
food , Polyphenolic antioxidants , Manganese(IV) , Chemiluminescence , Potassium permanganate , Luminol , Cerium(IV) , Flow injection analysis
Journal title :
Food Chemistry
Journal title :
Food Chemistry