Title of article
High sensitivity MOSFET-based neutron dosimetry
Author/Authors
Fragopoulou، نويسنده , , M. and Konstantakos، نويسنده , , V. and Zamani، نويسنده , , M. and Siskos، نويسنده , , S. and Laopoulos، نويسنده , , T. and Sarrabayrouse، نويسنده , , G.، نويسنده ,
Pages
4
From page
611
To page
614
Abstract
A new dosemeter based on a metal-oxide-semiconductor field effect transistor sensitive to both neutrons and gamma radiation was manufactured at LAAS-CNRS Laboratory, Toulouse, France. In order to be used for neutron dosimetry, a thin film of lithium fluoride was deposited on the surface of the gate of the device. The characteristics of the dosemeter, such as the dependence of its response to neutron dose and dose rate, were investigated. The studied dosemeter was very sensitive to gamma rays compared to other dosemeters proposed in the literature. Its response in thermal neutrons was found to be much higher than in fast neutrons and gamma rays.
Keywords
MOSFET’s , neutron dosimetry
Journal title
Astroparticle Physics
Record number
1991938
Link To Document