• Title of article

    Estimation of presampling modulation transfer function in synchrotron radiation microtomography

  • Author/Authors

    Mizutani، نويسنده , , Ryuta and Taguchi، نويسنده , , Keisuke and Takeuchi، نويسنده , , Akihisa and Uesugi، نويسنده , , Kentaro and Suzuki، نويسنده , , Yoshio، نويسنده ,

  • Pages
    5
  • From page
    615
  • To page
    619
  • Abstract
    The spatial resolution achieved by recent synchrotron radiation microtomographs should be estimated from the modulation transfer function (MTF) on the micrometer scale. Step response functions of a synchrotron radiation microtomograph were determined by the slanted edge method by using high-precision surfaces of diamond crystal and ion-milled aluminum wire. Tilted reconstruction was introduced to enable any edge to be used as the slanted edge by defining the reconstruction pixel matrix in an arbitrary orientation. MTFs were estimated from the step response functions of the slanted edges. The obtained MTFs coincided with MTF values estimated from square-wave patterns milled on the aluminum surface. Although X-ray refraction influences should be taken into account to evaluate MTFs, any flat surfaces with nanometer roughness can be used to determine the spatial resolutions of microtomographs.
  • Keywords
    Micro-CT , Line spread function , Test pattern , Focused ion-beam milling
  • Journal title
    Astroparticle Physics
  • Record number

    1991940