• Title of article

    Photomultipliers with microchannel plates

  • Author/Authors

    Anashin، نويسنده , , V.V. and Beschastnov، نويسنده , , P.M. and Golubev، نويسنده , , V.B. and Mironenko، نويسنده , , L.A. and Salnikov، نويسنده , , A.A. and Serednyakov، نويسنده , , S.I.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    7
  • From page
    103
  • To page
    109
  • Abstract
    In this paper we present the results of tests of photomultiplier tubes with microchannel plates (MCP—PMT), which were manufactured at INP as prototypes for future use in particle detectors. The MCP—PMTs have smaller height/diameter ratio in comparison with ordinary PMTs, a gain of 104–107 and can operate in a magnetic field. Considerable degradation of the tubes occurs at about 0.01 C/cm2 of collected anode charge. The PMT—MCPs show a clear single-electron peak. They were tested in setups with a quartz Cherenkov radiator and a NaI(TI) scintillation crystal.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1995
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    1993769