Title of article :
Coupling of the recoil mass spectrometer CAMEL to the γ-ray spectrometer GASP
Author/Authors :
Spolaore، نويسنده , , Annette P. and Ackermann، نويسنده , , D. and Bednarczyk، نويسنده , , P. and De Angelis، نويسنده , , G. and Napoli، نويسنده , , D. and Rossi Alvarez، نويسنده , , C. and Bazzacco، نويسنده , , D. and Burch، نويسنده , , R. and Müller، نويسنده , , L. and Segato، نويسنده , , G.F. and Scarlassara، نويسنده , , F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
6
From page :
500
To page :
505
Abstract :
A project has been realized to link the CAMEL recoil mass spectrometer to the GASP γ-spectrometer in order to perform high resolution and efficiency γ-recoil coincidence measurements. To preserve high flexibility and autonomy in the operation of the two complex apparatus a rough factor two of reduction in the overall heavy ion transmission was accepted in designing the optics of the particle transport from the GASP center to the CAMEL focal plane. The coupled configuration has been tested with the fusion reaction 58Ni (E = 212 MeV) + 64Ni, obtaining a mass resolution of 1300 and efficiency between ∼ 11% and ∼ 15% for different evaporation products.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1995
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
1994720
Link To Document :
بازگشت