Title of article
Comparison of charge collection in semiconductor detectors and timing resolution, using a sub-nanosecond transimpedance amplifier
Author/Authors
Rudge، نويسنده , , Alan، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
8
From page
169
To page
176
Abstract
A transimpedance amplifier, with a risetime of < 600 ps and a noise of < 1000 RMS electrons in a 500 MHz bandwidth, has been used for comparison of charge collection times in silicon, gallium arsenide and diamond detectors. The use of silicon detectors as trigger counters/hodoscopes is demonstrated, together with measured timing characteristics.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1995
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
1994858
Link To Document