Title of article
Position-sensitive MCP-based detectors with high timing resolution: some results and perspectives
Author/Authors
Feofilov، نويسنده , , G.A. and Stolyarov، نويسنده , , O.I. and Tsimbal، نويسنده , , F.A. and Valiev، نويسنده , , F.F. and Vinogradov، نويسنده , , L.I.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
6
From page
402
To page
407
Abstract
The microelectronic design and some results of experimental tests of three types of coordinate-sensitive Micro-Channel-Plane (MCP) based detectors with good timing properties are presented. The electrodynamic structures that could be used for the MCP fast signal readout (strips, delay lines and discrete pads) are always a compromise between the amplitude of the induced signal, propagation time delay, cross-talk and distortions of the signal at the output.
tests of these structures and experimental tests of the detectors done with the beams of MIPs and with the alpha-source, have shown a 130 μm spatial resolution and a 110 ps timing resolution obtained simultaneously for one detector.