Author/Authors :
Mochiki، نويسنده , , Koh-ichi and Uno، نويسنده , , Takayuki and Murata، نويسنده , , Yutaka and Kobayashi، نويسنده , , Hisao and Matsubayashi، نويسنده , , Masahito and Tsuruno، نويسنده , , Akira، نويسنده ,
Abstract :
A high-performance HARPICON tube has been equipped to an electronic imaging camera of Musashi Image Processing System for neutron radiography (NRG). A vertical limiting resolution of 1600 TV lines was obtained with an imaging size of 2048 × 2048 pixels using the EIAJ test chart-A. The camera was applied to neutron radiography at the NRG facility of JRR-3M at Japan Atomic Energy Research Institute and excellent neutron radiographs have been obtained. Durability of the HARPICON camera against neutron and gamma rays was proved.