• Title of article

    Energy and active volume calibration of Xe-filled proportional counters in the 1 to 10 KeV range

  • Author/Authors

    Schanda، نويسنده , , Ulrich and Urban، نويسنده , , Andreas، نويسنده ,

  • Pages
    7
  • From page
    79
  • To page
    85
  • Abstract
    Proportional counters (PC), containing xenon as a component of the counting gas, can be calibrated in the energy range of 1 to 10 keV by means of an external cerium-X-ray source. Three composite peaks emerge at 1.1, 5.1 and 9.8 keV. Additionally the gas gain along the anode wire can be scanned by collimating the cerium-X-ray source, thus the actual active volume of the PC can be determined. This calibration method is used in the GALLEX solar neutrino experiment.
  • Journal title
    Astroparticle Physics
  • Record number

    1999398