Author/Authors :
Henneck، نويسنده , , R. and Bialkowski، نويسنده , , J. and Bühler، نويسنده , , P. San Martini، نويسنده , , R. and Mchedlishvili، نويسنده , , A. T. Zehnder، نويسنده , , A. and Siegmund، نويسنده , , O.، نويسنده ,
Abstract :
The VUV (30–1200Å) detector described consists of a compact ‘Micro-Channel-Plate-Wedge-and-Strip anode’ imaging system and of the corresponding data read-out electronics with 32 Mbyte of memory. The performance of this space qualified detector system was measured in the FUV. We obtained a spatial resolution of about 70 μm over the full detector area of ∼7 cm2. Image distortions are small and can easily be corrected for. Detector efficiency depends on the choice of photo-cathode and ranges between 50% in the FUV up to 90% in the EUV.