Title of article :
Low dose-rate irradiation set-up for scintillating crystals
Author/Authors :
DʹAmbrosio، نويسنده , , C and Ercoli، نويسنده , , C and Jaaskelainen، نويسنده , , S and Rosso، نويسنده , , E and Wicht، نويسنده , , P، نويسنده ,
Pages :
8
From page :
119
To page :
126
Abstract :
With a 1 Ci60Co-source we measured the radiation damage and its recovery in PbWO4 crystals. The method consists in measuring the anode current produced in the silicon diode of a Hybrid PhotoMultiplier Tube (HPMT) during the irradiation of the crystal. At regular time intervals, we took photon counting spectra with the same HPMT. By checking the resulting channel positions of the photoelectron peaks, we controlled the stability of our experimental set-up. Temperature control and monitoring were achieved by forced thermostatized water circulation inside the irradiation box and by using a PT100 and two AD590 temperature probes. First results on irradiations of a few PbWO4 crystals are presented and future plans are communicated.
Journal title :
Astroparticle Physics
Record number :
2001049
Link To Document :