Title of article :
A simplified read out for high-resolution linear gas detector
Author/Authors :
Bordessoule، نويسنده , , Michel and Bartol، نويسنده , , François A. Lemonnier، نويسنده , , Marc and Santiard، نويسنده , , Jean-Claude، نويسنده ,
Pages :
7
From page :
79
To page :
85
Abstract :
We present here measurements on a wire-pad gas detector. The detector resolution is derived from the charge profile and the electronic noise. All data are compared to a model. The useful length of the detector is 8 cm and the measured FWHM mean-spatial resolution is around 110 μm. The electronics relies essentially on a single chip, the GASSIPLEX, connected to a low-end ADC and PC, the later two elements defining the acquisition speed.
Keywords :
Analogue read out , centroid , Proportional counter , Gas detector , Linear detector
Journal title :
Astroparticle Physics
Record number :
2001316