Abstract :
It is shown that fine geometric modelling of the HPGe detector structure using the GEANT system leads to a good match between the calculated and measured spatial dependence of efficiency for point sources in the region of photon energies from 60 keV to 1115 keV.
Keywords :
?-Ray spectroscopy , HPGe detectors , Monte-Carlo calculation