Author/Authors :
Zavattini، نويسنده , , G. and Feroci، نويسنده , , M. and Budtz-Jّrgensen، نويسنده , , C. and Christensen، نويسنده , , F. and Frontera، نويسنده , , F. and Gambaccini، نويسنده , , M. and Ortuٌo-Prados، نويسنده , , F. and Pareschi، نويسنده , , G.، نويسنده ,
Abstract :
A prototype Xe ÷ 10% CH4 microstrip detector was used to study the K-edge discontinuity in the pulse-height distribution as a function of the energy of incident X-rays. The electronics used was such that a pulse-shape rejection could be made of K-fluorescence reabsorption in the detector. The measured jump at the K-edge was (169 ± 12) eV. It was also seen that Kα- and Kβ-decay channels produce different pulse heights for the same deposited energy. We find that the difference in energy between the calibration lines of the Kβ2- and the Kβ1.3-escape peaks is (93 ± 12) eV whereas between the Kβ1.3 and the Kα-escape peaksʹ calibration the energy gap is (69 ± 7) eV.
Keywords :
Astronomical and Space Research Instrumentation , Radiation Spectrometers and Spectroscopic Techniques , Radiation detectors