Title of article :
Preparation of a beam quality indicator for effective energy determinations of continuum beams: establishment of traceability
Author/Authors :
Matsubayashi، نويسنده , , Masahito and T. Lindsay، نويسنده , , John and Kobayashi، نويسنده , , Hisao، نويسنده ,
Pages :
7
From page :
165
To page :
171
Abstract :
A new beam quality indicator (BQI) was designed and fabricated to determine effective energies of beams extracted from neutron radiography facilities. Performances of the five new BQIs were compared with the original BQI which was recently proposed and tested by various beams. Non-filtered thermal neutrons, filtered thermal neutrons, and cold neutrons from a guide tube were used in the performance test program. The new BQIs were also examined by four different detection systems using a combination of a Gd converter and a X-ray film, a neutron imaging plate, a cooled charge coupled device camera, and a silicon intensified target tube camera.
Keywords :
Effective energy , Comparison of BQIs , Traceability , Comparison of detection systems , Filtered neutron beams , Continuum neutron beam
Journal title :
Astroparticle Physics
Record number :
2008591
Link To Document :
بازگشت