Author/Authors :
Pugliesi، نويسنده , , R and Geraldo، نويسنده , , L.P and Andrade، نويسنده , , M.L.G and M.O.Menezes and Pereira، نويسنده , , M.A.S and Maizato، نويسنده , , M.J.S، نويسنده ,
Abstract :
The neutron radiography technique was employed to inspect an artificial heart prototype which is being developed to provide blood circulation for patients expecting heart transplant surgery. The radiographs have been obtained by the direct method with a gadolinium converter screen along with the double coated Kodak-AA emulsion film. The artificial heart consists of a flexible plastic membrane located inside a welded metallic cavity, which is employed for blood pumping purposes. The main objective of the present inspection was to identify possible damages in this plastic membrane, produced during the welding process of the metallic cavity. The obtained radiographs were digitized as well as analysed in a PC and the improved images clearly identify several damages in the plastic membrane, suggesting changes in the welding process.