Title of article :
Coincidence corrections for a multi-detector gamma spectrometer
Author/Authors :
Britton، نويسنده , , R. and Burnett، نويسنده , , J.L. and Davies، نويسنده , , A.V. and Regan، نويسنده , , P.H.، نويسنده ,
Pages :
6
From page :
20
To page :
25
Abstract :
List-mode data acquisition has been utilised in conjunction with a high-efficiency γ–γ coincidence system, allowing both the energetic and temporal information to be retained for each recorded event. Collected data is re-processed multiple times to extract any coincidence information from the γ-spectroscopy system, correct for the time-walk of low-energy events, and remove accidental coincidences from the projected coincidence spectra. The time-walk correction has resulted in a reduction in the width of the coincidence delay gate of 18.4±0.4%, and thus an equivalent removal of ‘background’ coincidences. The correction factors applied to ~5.6% of events up to ~500 keV for a combined 137Cs and 60Co source, and are crucial for accurate coincidence measurements of low-energy events that may otherwise be missed by a standard delay gate. By extracting both the delay gate and a representative ‘background’ region for the coincidences, a coincidence background subtracted spectrum is projected from the coincidence matrix, which effectively removes ~100% of the accidental coincidences (up to 16.6±0.7% of the total coincidence events seen during this work). This accidental-coincidence removal is crucial for accurate characterisation of the events seen in coincidence systems, as without this correction false coincidence signatures may be incorrectly interpreted.
Keywords :
Coincidence systems , Accidental coincidence , gamma spectroscopy , Time-walk
Journal title :
Astroparticle Physics
Record number :
2009651
Link To Document :
بازگشت