Title of article :
Development of CdTe pixel detector for synchrotron radiation experiments
Author/Authors :
Hirono، نويسنده , , T. and Toyokawa، نويسنده , , H. Kawano-Furukawa، نويسنده , , Y. and Kawase، نويسنده , , M. and Ohata، نويسنده , , T. and Wu، نويسنده , , S. and Ikeda، نويسنده , , H. and Sato، نويسنده , , G. and Takahashi، نويسنده , , T. and Watanabe، نويسنده , , S.، نويسنده ,
Pages :
4
From page :
64
To page :
67
Abstract :
This study describes the development of a CdTe pixel detector for high-energy X-ray diffraction experiments at synchrotron radiation facilities. We have developed three prototypes. The third prototype detector (SP8-02B) was designed with a pixel size of 200 μ m × 200 μ m and a matrix of 20×50 pixels. The sensor was bonded to an application-specific integrated circuit with a preamplifier, shaper, window-type discriminator, and a 20-bit counter for each pixel. The SP8-02B detector improved the uniformity between pixels in one chip in comparison with the previous prototypes. The long-term stability was measured at various temperatures. A moderate cooling achieved good stability that was not affected by the polarization effect.
Keywords :
CdTe , Readout ASIC , Photon counting , Pixel detector , x-ray imaging