Katayama، نويسنده , , H and Shouho، نويسنده , , M and Kohmura، نويسنده , , T and Katayama، نويسنده , , K and Yoshita، نويسنده , , K and Tsunemi، نويسنده , , H and Kitamoto، نويسنده , , S and Hayashida، نويسنده , , K and Miyata، نويسنده , , K. Hashimotodani، نويسنده , , K and Koyama، نويسنده , , K and Ricker، نويسنده , , G and Bautz، نويسنده , , M.W and Foster، نويسنده , , R and Kissel، نويسنده , , S، نويسنده ,
Abstract :
We measured the optical and the X-ray transmission of the optical blocking filters for the X-ray Imaging Spectrometers (XISs) which are the X-ray CCD cameras of the ASTRO-E satellite. We conclude that the oxidation of the aluminum reduces the optical transmission down to ∼60–70% of the theoretical value of the aluminum. We achieved optical transmission below 5×10−5 in the range from 4000 to 9500 Å by using aluminum thickness of 1200 Å, while the theoretical calculation requires 800 Å. The measurement of absolute quantum efficiency of XIS is also performed at several particular energies. We confirmed 20% quantum efficiency at 0.5 keV for the XIS engineering model (XIS EM).
Keywords :
CCD , Calibration , ASTRO-E , Quantum efficiency