• Title of article

    Detector embedded device for continuous reset of charge amplifiers: choice between bipolar and MOS transistor

  • Author/Authors

    Guazzoni، نويسنده , , C. and Sampietro، نويسنده , , M. and Fazzi، نويسنده , , A.، نويسنده ,

  • Pages
    4
  • From page
    447
  • To page
    450
  • Abstract
    This work discusses the criteria for the choice of the reset device in a charge amplifier whose front-end is integrated on high-resolution semiconductor detectors. The performances achievable using a bipolar or a MOS transistor as reset device are compared in terms of the linearity of the response and of the added noise as a function of the detector leakage current. The additional constraints in term of available technology and layout compatibility show the advantages of MOS transistors operated in sub-threshold mode.
  • Keywords
    Front-end electronics , Charge amplifiers , X-ray detectors , X-ray spectroscopy
  • Journal title
    Astroparticle Physics
  • Record number

    2011644