Author/Authors :
Chervenak، نويسنده , , J.A and Grossman، نويسنده , , E.N and Irwin، نويسنده , , K.D and Martinis، نويسنده , , John M and Reintsema، نويسنده , , C.D and Allen، نويسنده , , C.A and Bergman، نويسنده , , D.I and Moseley، نويسنده , , S.H and Shafer، نويسنده , , R، نويسنده ,
Abstract :
We report on the implementation of a multiplexer that uses superconducting quantum interference devices (SQUIDs) to read out low-impedance cryogenic detectors. Using prototype chips, a circuit was built which interfaces eight input SQUID channels with a close-packed array of eight transition-edge sensor (TES) infrared bolometers. Circuit elements were measured and crosstalk specifications are reported. Digital feedback is employed to flux-lock a single element in the array of SQUIDs.