Title of article :
A position-sensitive tubular scintillator-based detector as an alternative to a 3He-gas-based detector for neutron-scattering instruments
Author/Authors :
Nakamura، نويسنده , , T. and Katagiri، نويسنده , , M. H. Toh، نويسنده , , K. and Honda، نويسنده , , K. and Suzuki، نويسنده , , H. and Ebine، نويسنده , , M. and Birumachi، نويسنده , , A. and Sakasai، نويسنده , , K. and Soyama، نويسنده , , K.، نويسنده ,
Pages :
5
From page :
42
To page :
46
Abstract :
A position-sensitive tubular scintillator-based neutron detector is proposed as an alternative to a 3He-gas-based detector. The detector has a neutron-detecting element constructed from rolled ZnS/6LiF scintillator screens that sandwich wavelength-shifting (WLS) fibre coils (SFC element). Multiple SFC elements are enclosed in an aluminium tube in a row to form a one-dimensional position-sensitive neutron detector. The design of the WLS fibre coil, which was determined by performing basic experiments, comprised two 0.75-mm-diameter WLS fibres wound in parallel at a pitch of 1.5 mm. A 64-element detector with a pixel size of 22 mm×20 mm (width×length) successfully demonstrated the detection principle. The tubular shape of the new detector is similar to the usual 25-mm-diameter 3He tube, making this an alternative detector with the potential to be installed in a vacuum tank for inelastic-neutron-scattering instruments.
Keywords :
3He-gas-based detector , Scintillator-based neutron detector , Wavelength-shifting fibre , Position-sensitive detector , Aluminium tube , Vacuum
Journal title :
Astroparticle Physics
Record number :
2011706
Link To Document :
بازگشت