Title of article
Measurement of the γ-ray energy resolution function of EJ301 liquid scintillator using a dual channel ADC
Author/Authors
Ivan Cecconello، نويسنده , , M. and Donato، نويسنده , , M. and Marini-Bettolo، نويسنده , , C. and Conroy، نويسنده , , S. and Sangaroon، نويسنده , , S. and Ericsson، نويسنده , , G.، نويسنده ,
Pages
4
From page
34
To page
37
Abstract
A comparison of three different methods for the energy calibration of liquid scintillators using γ-ray sources using only a delay unit and a dual channel ADC is presented. Single Compton edge measurements combined with MCNP simulations of the pulse height spectra are compared with the results obtained using Compton coincidence methods. Good agreement between the three methods is found. Measurements from the three different methods are then combined to provide the best estimate of the energy resolution function. The technique here presented could be easily used in laboratory courses as high performing ADCs become more common and affordable.
Keywords
Liquid scintillator , Coincidence technique , Energy resolution function , Compton edge
Journal title
Astroparticle Physics
Record number
2012218
Link To Document