Author/Authors :
Hennig، نويسنده , , A. and Fransen، نويسنده , , C. and Hennig، نويسنده , , W. and Pascovici، نويسنده , , Martin G. and Warr، نويسنده , , N. and Weinert، نويسنده , , M. and Zilges، نويسنده , , A.، نويسنده ,
Abstract :
The influence on γ-ray spectra of differential nonlinearities (DNL) in subranging, pipelined analog-to-digital converts (ADCs) used for digital γ-ray spectroscopy was investigated. The influence of the DNL error on the γ-ray spectra, depending on the input count-rate and the dynamic range has been investigated systematically. It turned out that the DNL becomes more significant in γ-ray spectra with larger dynamic range of the spectroscopy system. An event-by-event offline correction algorithm was developed and tested extensively. This correction algorithm works especially well for high dynamic ranges.
Keywords :
?-Ray spectroscopy , HPGe detectors , Differential nonlinearities , Analog-to-digital conversion , Digital signal-processing