Xu، نويسنده , , Cheng and Chen، نويسنده , , Han and Persson، نويسنده , , Mats and Karlsson، نويسنده , , Staffan and Danielsson، نويسنده , , Mats and Svensson، نويسنده , , Christer and Bornefalk، نويسنده , , Hans، نويسنده ,
Abstract :
We investigated the energy resolution of a segmented silicon strip detector for photon-counting spectral computed tomography (CT). The detector response to different monochromatic photon energies and various photon fluxes was characterized at the Elettra synchrotron. An RMS energy resolution of 1.50 keV has been demonstrated for 22 keV photons at zero flux, and it deteriorated as a function of input count rate at a rate of 5.1 eV mm2/Mcps. The charge sharing effect has been evaluated. The results show that around 11.1% of the interacting photons experience charge sharing for 22 keV photons and 15.3% for 30 keV.
Keywords :
Photon counting , Synchrotron , Energy resolution , Silicon strip detector , Spectral computed tomography