Title of article :
Sparks in MSGC and GEM detectors and a robust alternative
Author/Authors :
B. Baruth، نويسنده , , O and Keller، نويسنده , , S and Werthenbach، نويسنده , , U and Zech، نويسنده , , G and Zeuner، نويسنده , , T، نويسنده ,
Pages :
4
From page :
272
To page :
275
Abstract :
Sparks in MSGCs can be induced by strongly ionizing particles that damage the anodes. In MSGCs combined with GEMs, the MSGC gain can be reduced below the sparking threshold. However, occasionally discharges in the GEM can produce large secondary sparks at the MSGC. We discuss the dependence of the spark rate on geometrical and electrical parameters of the detectors. Different gas mixtures are compared. The most efficient way to avoid secondary sparks is to reduce the electric field between GEM and MSGC to below 6.6 kV/cm. We present first results from a robust alternative to MSGCs, the groove detector.
Keywords :
Secondary sparks , Groove chamber , GEM sparks
Journal title :
Astroparticle Physics
Record number :
2013367
Link To Document :
بازگشت