Title of article :
Studying the full energy peak efficiency for a two γ-detectors combination of different dimensions
Author/Authors :
El-Khatib، نويسنده , , Ahmed M. and Krar، نويسنده , , Mohamed E. and Badawi، نويسنده , , Mohamed S.، نويسنده ,
Pages :
7
From page :
50
To page :
56
Abstract :
The present work investigates the validity of extending the efficiency transfer technique (ET) to accommodate detection system composed of more than one detector [two NaI (Tl) detectors]. An arbitrary located radioactive point and volumetric cylindrical sources [152Eu aqueous radioactive sources covering the energy range from 121 keV up to 1408 keV] were used in order to get the full-energy peak efficiency (FEPE) based on the effective solid angle ratio. In addition, self attenuation of source matrix, the attenuation by the source container and the detector housing materials are also treated. Results are compared with those measured by a combination of two NaI(Tl) detectors of different volumes (3″×3″ and 2″×2″) with resolution (FWHM) at 662 keV equal to 7.5% and 8.5% respectively. The comparison between the calculated and the measured full-energy peak efficiency values reveal a good agreement.
Keywords :
Two ?-detector combination , Efficiency transfer technique (ET) , Effective solid angle , Full-energy peak efficiency (FEPE)
Journal title :
Astroparticle Physics
Record number :
2013554