Title of article :
Medical X-ray imaging with energy windowing
Author/Authors :
Lindner، نويسنده , , M. and Blanquart، نويسنده , , L. and Fischer، نويسنده , , P. and Krüger، نويسنده , , H. and Wermes، نويسنده , , N.، نويسنده ,
Pages :
6
From page :
229
To page :
234
Abstract :
A photon counting pixel read-out chip for medical X-ray imaging with energy windowing is presented. It consists of 32×32 pixels of 200 μm×200 μm size, each of them contains two discriminators and two 18-bit counters. Both discriminator thresholds can be adjusted individually for good image homogeneity and a circuit has been designed to avoid threshold drift. Measurements on the noise behaviour, threshold variation and threshold drift are shown.
Keywords :
pixel , Energy Window , x-ray imaging , Photon counting
Journal title :
Astroparticle Physics
Record number :
2015989
Link To Document :