Author/Authors :
Friedrich، نويسنده , , S. and Niedermayr، نويسنده , , T. and Drury، نويسنده , , O. and Cunningham، نويسنده , , M.F. and van den Berg، نويسنده , , M.L. and Ullom، نويسنده , , J.N. and Loshak، نويسنده , , A. and Funk، نويسنده , , Stephen T. and Cramer، نويسنده , , S.P. and Batteux، نويسنده , , J.D. and See، نويسنده , , E. and Frank، نويسنده , , M. and Labov، نويسنده , , S.E.، نويسنده ,
Abstract :
We have built a two-stage adiabatic demagnetization refrigerator (ADR) to operate cryogenic high-resolution X-ray detectors in synchrotron-based fluorescence applications. The detector is held at the end of a 40 cm cold finger that extends into a UHV sample chamber. The ADR attains a base temperature below 100 mK with about 20 h hold time below 400 mK, and does not require pumping on the liquid He bath. We will discuss cryostat design and performance.