Title of article :
Efficiency calibration of a HPGe detector in the [46.54–2000] keV energy range for the measurement of environmental samples
Author/Authors :
Daza، نويسنده , , M.J. and Quintana، نويسنده , , B. and Garc??a-Talavera، نويسنده , , M. and Fern?ndez، نويسنده , , F.، نويسنده ,
Abstract :
Here we describe a general method of calibrating the efficiency of a Ge γ-ray spectrometer. The method, which is based on the work of Quintana and Fernández (Appl. Radiat. Isot. 47 (1996) 911), can now be applied to many different experimental set-ups including both liquid and solid environmental samples. The method requires two different types of experimental inputs. Firstly, it requires measurements with radioactive sources emitting cascades of γ rays covering the energy range of interest. Secondly, it requires measurements with sources emitting isolated γ rays in order to provide coincidence-summing corrections. On this basis, we establish a general function to describe the energy dependence of the efficiency for the particular geometry and source matrix. The method has been applied to 11 different experimental arrangements to provide efficiency calibrations over the range 46.54–2000 keV with associated uncertainties ranging from 0.1% to 1.8%. This allows high precision measurements with environmental samples, which often have very low activities.
Keywords :
?-ray spectrometry , Extended sources , Coincidence summing , efficiency calibration
Journal title :
Astroparticle Physics