Title of article :
LaCl3:Ce coincidence signatures to calibrate gamma-ray detectors
Author/Authors :
McIntyre، نويسنده , , Justin I. and Schrom، نويسنده , , Brian T. and Cooper، نويسنده , , Mathew W. and Day، نويسنده , , Anthony and Hayes، نويسنده , , James C. and Heimbigner، نويسنده , , Tom R. and Hubbard، نويسنده , , Charles W. and Ripplinger، نويسنده , , Michael D. and Suarez، نويسنده , , Reynold، نويسنده ,
Pages :
4
From page :
201
To page :
204
Abstract :
Calibrating the gamma-ray detection efficiency of radiation detectors in a field environment is difficult under most circumstances. To counter this problem we have developed a technique that uses a Cerium doped Lanthanum-Tri-Chloride (LaCl3:Ce) scintillation detector to provide gated gammas (Guillot-Noël et al. (1999) [1] and van Loef et al. (2001) [2]). Exploiting the inherent radioactivity of the LaCl3:Ce due to the long-lived radioactive isotope 138La (t1/2=1.06×1011 yr) allows the use of the 788 and 1436 keV gammas as a measure of efficiency. In this paper, we explore the effectiveness of using the beta-gamma coincidences radiation LaCl3:Ce detector to calibrate the energy and efficiency of a number of gamma-ray detectors.
Keywords :
Lanthanum-Tri-Chloride , radiation detector , Calibration , Coincidence techniques
Journal title :
Astroparticle Physics
Record number :
2017439
Link To Document :
بازگشت