Title of article :
Simulation and characterization of different setups for gamma ray detection using SiPMs and LYSO scintillators
Author/Authors :
Benetti، نويسنده , , M. and Tarolli، نويسنده , , A. and Giacomini، نويسنده , , G. and Piemonte، نويسنده , , C. and Dalla Betta، نويسنده , , G.-F.، نويسنده ,
Pages :
5
From page :
61
To page :
65
Abstract :
Silicon photomultipliers (SiPMs) coupled to fast bright scintillators, like cerium doped silicate based crystals, allow the construction of compact gamma ray detectors. In this paper we discuss simulation results obtained from Monte Carlo ray tracing tools applied to SiPM and LYSO systems. We address the importance of three key factors in light propagation: the scintillator wrapping, the coupling medium, and the detector coating. We also propose a simple experiment to verify some of the findings related to the investigation of diffusive wrappings.
Keywords :
SiPM , Gamma ray detectors , Monte Carlo optical simulation , scintillator
Journal title :
Astroparticle Physics
Record number :
2018397
Link To Document :
بازگشت