Title of article :
Secondary radiation in transmission-type X-ray tubes: Simulation, practical issues and solution in the context of X-ray microtomography
Author/Authors :
Boone، نويسنده , , Matthieu N. and Vlassenbroeck، نويسنده , , Jelle and Peetermans، نويسنده , , Steven and Van Loo، نويسنده , , Denis and Dierick، نويسنده , , Manuel and Van Hoorebeke، نويسنده , , Luc، نويسنده ,
Abstract :
In laboratory-based X-ray radiography and computed tomography, the X-rays are assumed to originate from one single focal spot with a finite spot size, which is generated by focussing accelerated electrons on the target material. However, apart from this focal spot, X-rays can also be produced elsewhere in the tube system. A major contribution of this parasitic radiation originates from electrons that are backscattered from the target material, into the X-ray tube system, where they can produce so-called off-focus or secondary X-rays. This phenomenon has been widely studied for rotating anode X-ray tubes in medical imaging systems, but not for transmission-type microfocus X-ray tubes. This paper presents a study on the origin of secondary radiation in this kind of X-ray tubes, which is performed by Monte Carlo simulations and by experimental measurements. The impact of this phenomenon on the imaging process is studied, and two correction methods are proposed, both on the hardware and on the software levels.
Keywords :
X-ray tube , Focal spot , Monte Carlo simulation , Off-focal radiation , Secondary radiation , BEAMnrc
Journal title :
Astroparticle Physics