Author/Authors :
Agosteo، نويسنده , , S. and Borsato، نويسنده , , E. and Dal Corso، نويسنده , , F. and Fazzi، نويسنده , , A. and Gonella، نويسنده , , F. and Introini، نويسنده , , M.V. and Lippi، نويسنده , , I. and Lorenzoli، نويسنده , , M. and Modenese، نويسنده , , L. and Montecassiano، نويسنده , , F. and Pegoraro، نويسنده , , M. and Pola، نويسنده , , A. and Varoli، نويسنده , , V. and Zotto، نويسنده , , P.، نويسنده ,
Abstract :
A Proton Irradiation Chamber aiming to perform radiation tests of electronic components was developed. The precision on the measurement of the ion currents was pushed beyond the resolution of the picoammeter by means of a series of collimators on the beam showing a linear correlation among the currents measured on them and the smaller, not measurable, current on the target. As an example of the obtained results the tests done on a Si microdosimeter and a power p-MOS are reported.