An automatic energy-calibration method for segmented germanium detectors
Author/Authors
Hu، نويسنده , , Z. and Glasmacher، نويسنده , , T. and Mueller، نويسنده , , W.F. and Wiedenhِver، نويسنده , , I.، نويسنده ,
Pages
5
From page
715
To page
719
Abstract
An automatic gain-matching method has been developed for the energy calibrations of the 32 segments of a large-volume germanium detector. The method calibrates the small-volume segments relative to the central contact, which collects the charge from the total germanium crystal volume and can be efficiently calibrated with known sources. The method presented here can employ photons of arbitrary energy. With the use of in-beam γ-rays for segment calibrations it presents an efficient way to perform continuous energy calibrations in experiments.
Keywords
Germanium detectors , Gain matching , Segmented detectors