Author/Authors :
Dinca، نويسنده , , L.E and Dorenbos، نويسنده , , P and de Haas، نويسنده , , J.T.M and Bom، نويسنده , , V.R and van Eijk، نويسنده , , C.W.E، نويسنده ,
Abstract :
Some scintillating materials offer the possibility of measuring well separated α and γ scintillation response using a single crystal. Eventually aiming at thermal neutron detection using 6Li or 10B admixture, pulse shape discrimination measurements were made on three scintillators: CsI:Tl, CsI:Na and pure BaF2 crystals. A very good α/γ discrimination was obtained using 22Na, 241Am (γ) and 244Cm (α) radioactive sources.