Characterization of the Hamamatsu R11410-10 3-in. photomultiplier tube for liquid xenon dark matter direct detection experiments
Author/Authors
Lung، نويسنده , , K. and Arisaka، نويسنده , , K. and Bargetzi، نويسنده , , A. and Beltrame، نويسنده , , Joseph P. and Cahill، نويسنده , , A. and Genma، نويسنده , , T. and Ghag، نويسنده , , C. and Gordon، نويسنده , , D. and Sainz، نويسنده , , J. and Teymourian، نويسنده , , A. and Yoshizawa، نويسنده , , Y.، نويسنده ,
Pages
8
From page
32
To page
39
Abstract
To satisfy the requirements of the next generation of dark matter detectors based on the dual phase TPC, Hamamatsu, in close collaboration with UCLA, has developed the R11410-10 photomultiplier tube. In this work, we present the detailed tests performed on this device. High QE ( ∼ 30 % ) accompanied by a low dark count rate (50 Hz at 0.3 PE) and high gain (1×107) with good single PE resolution have been observed. A comprehensive screening measurement campaign is ongoing while the manufacturer quotes a radioactivity of 20 mBq/PMT. These characteristics show the R11410-10 to be particularly suitable for the forthcoming zero background liquid xenon detectors.
Keywords
Photomultiplier tubes , Astroparticle physics , Dark matter