Title of article
Determination of a source in a box with two detectors. I. non-absorbing media
Author/Authors
Presler، نويسنده , , O. and Pelled، نويسنده , , O. and German، نويسنده , , U. and Leichter، نويسنده , , Y. and Alfassi، نويسنده , , Z.B.، نويسنده ,
Pages
12
From page
314
To page
325
Abstract
It was found that the activity and the position of a radioactive point source (hot spot) in a large box containing low absorbing material can be measured accurately using two NaI(Tl) gamma-ray detectors at the opposite sides of the box i.e. at 180° one to another. The harmonic mean of the count rates square roots measured with the two detectors was found to be independent of the hot spot position and depends only on the source activity, the box size and the gamma-rays energy. The activity of the radioactive point source can be calculated from the above-mentioned mean. The position of the point source can be calculated from the ratio of the two detectors count rates.
Keywords
Virtual point detectors , A source in a box , Unknown radioactive sources
Journal title
Astroparticle Physics
Record number
2020541
Link To Document