Author/Authors :
Chen، نويسنده , , J.B. and Zheng، نويسنده , , Z.J. and Peng، نويسنده , , H.S. and Tang، نويسنده , , C.H. and Zhang، نويسنده , , B.H. and Ding، نويسنده , , Y.K. and Chen، نويسنده , , M. and Chen، نويسنده , , H.S. and Li، نويسنده , , C.G. and Wen، نويسنده , , T.S and Yu، نويسنده , , R.Z.، نويسنده ,
Abstract :
The time response functions of an ultrafast quenched plastic scintillation detector used to measure neutron time of flight spectra were calibrated by utilizing cosmic rays and implosion neutrons from DT-filled capsules at the Shenguang II laser facility. These sources could be regarded as δ function pulses due to their much narrower time widths than those of the time response functions of the detection system. The results showed that the detector responses to DT neutrons and to cosmic rays were 1.18 and 0.96 ns FWHM, respectively.
Keywords :
Ultrafast quenched plastic scintillation detector , Laser facility , ? function pulsed neutron source , Cosmic ray , Time response function