Author/Authors :
Farahmand، نويسنده , , M. and Bos، نويسنده , , A.J.J. and Huizenga، نويسنده , , J. and De Nardo، نويسنده , , L. and van Eijk، نويسنده , , C.W.E.، نويسنده ,
Abstract :
By employing a Gas Electron Multiplier a new type of mini multi-element Tissue-Equivalent Proportional Counter (TEPC) has been designed and constructed. In this paper, we describe the design of this novel counter. The first pulse height measurements with this counter for both methane- and propane-based Tissue Equivalent gases are presented. These results show promising properties for application of this novel type TEPC in microdosimetric measurements.
Keywords :
microdosimetry , TEPC , Proportional counter , GEM