Title of article :
The high resolution gamma imager (HRGI): a CCD based camera for medical imaging
Author/Authors :
Lees، نويسنده , , John.E. and Fraser، نويسنده , , George.W. and Keay، نويسنده , , Adam and Bassford، نويسنده , , David and Ott، نويسنده , , Robert and Ryder، نويسنده , , William، نويسنده ,
Pages :
4
From page :
23
To page :
26
Abstract :
We describe the High Resolution Gamma Imager (HRGI): a Charge Coupled Device (CCD) based camera for imaging small volumes of radionuclide uptake in tissues. The HRGI is a collimated, scintillator-coated, low cost, high performance imager using low noise CCDs that will complement whole-body imaging Gamma Cameras in nuclear medicine. Using 59.5 keV radiation from a 241Am source we have measured the spatial resolution and relative efficiency of test CCDs from E2V Technologies (formerly EEV Ltd.) coated with Gadox (Gd2O2S(Tb)) layers of varying thicknesses. The spatial resolution degrades from 0.44 to 0.6 mm and the detection efficiency increases (×3) as the scintillator thickness increases from 100 to 500 μm. We also describe our first image using the clinically important isotope 99mTc. The final HRGI will have intrinsic sub-mm spatial resolution (∼0.7 mm) and good energy resolution over the energy range 30–160 keV.
Keywords :
CCDs , MEDICAL IMAGING , Gadox , 241Am
Journal title :
Astroparticle Physics
Record number :
2021458
Link To Document :
بازگشت