Title of article :
The new effective detector for digital scanning radiography
Author/Authors :
Babichev، نويسنده , , E.A. and Baru، نويسنده , , S.E. and Groshev، نويسنده , , V.R. and Khabakhpashev، نويسنده , , A.G. and Leonov، نويسنده , , V.V. and Neustroev، نويسنده , , V.A. and Porosev، نويسنده , , V.V. and Savinov، نويسنده , , G.A. and Shekhtman، نويسنده , , L.I.، نويسنده ,
Pages :
4
From page :
57
To page :
60
Abstract :
The new effective one-dimensional detector for digital scanning radiography is presented. The progress in the design of low-noise electronics allowed the Multistrip Ionization Chamber to be used as the X-ray detector. The detector has been designed for use in the Low-dose Digital Radiographic Device “Siberia” instead of a multiwire proportional chamber. The working gas is Xe under 12 atm pressure. The main detector parameters are as follows: total number of channels, −1024; channel size, −0.4×0.4 mm2; space resolution, −1.25 lp/mm; contrast sensitivity, −1%; dynamic range, −480; surface dose for chest image, −3 to 5 mR.
Keywords :
X-ray detector , ionization chamber , digital radiography
Journal title :
Astroparticle Physics
Record number :
2021488
Link To Document :