• Title of article

    “Scintillator-photodiode” detectors for scanning introscopy with high-spatial resolution

  • Author/Authors

    Ryzhikov، نويسنده , , V. and Kozin، نويسنده , , D. and Grinyov، نويسنده , , B. and Lysetskaya، نويسنده , , E. and Katrunov، نويسنده , , K. and Kvyatnitskaya، نويسنده , , V. and Chernikov، نويسنده , , V.، نويسنده ,

  • Pages
    5
  • From page
    58
  • To page
    62
  • Abstract
    The improvement of spatial resolution, together with reducing the dose load upon the tested object and making the scanning time shorter, is an essential aim in designing new radiographic systems. In this paper, we consider a possibility to use “scintillator-photodiode” detectors in radiographic systems in order to obtain resolution better than 1 line pair per mm (detector array pitch less than 1 mm). A possibility is considered to use various types of scintillators (CsI(Tl), CdWO4, ZnSe(Te)), various kinds of scintillation elements (separate scintillation elements, scintillation monoplates, plates made of a small-crystalline scintillator and a binding substance) with differently treated surfaces, different binders, reflective coatings, photoconverters (photodiode or photodiode+amplifier+commutator in one housing). It has been shown that ZnSe(Te) scintillators are preferable for detectors of X-ray and low-energy gamma-radiation (20–100 keV) with array pitch as small as 0.2 mm. CsI(Tl) should be used for 70–200 keV energies (0.4 mm pitch), CdWO4 crystals—for 0.3–10 MeV energies (pitch 0.4 mm).
  • Keywords
    Detectors , Spatial resolution , X-Ray , Sensitivity , Scintillators
  • Journal title
    Astroparticle Physics
  • Record number

    2021754