Title of article :
Computer simulation of time-of-flight mass spectrometers: calculations of mass spectra and spatial distributions of ions
Author/Authors :
Veryovkin، نويسنده , , Igor V. and Chen، نويسنده , , Chun-Yen and Calaway، نويسنده , , Wallis F. and Pellin، نويسنده , , Michael J. and Lee، نويسنده , , Typhoon Lee، نويسنده ,
Pages :
8
From page :
345
To page :
352
Abstract :
In the present work, an advanced algorithm to simulate complex ion optical systems is described, and results of its application to time-of-flight mass spectrometers with laser post-ionization of secondary neutrals are presented. To examine ion optics designs for such instruments and to find optimized solutions, a combination of charged particle optics modeling software (for simulating large sets of ion trajectories) and database software (for processing results of the simulations with database management routines) was introduced and used to aid the design of new instruments developed at Argonne National Laboratory.
Keywords :
Ion Optics , time-of-flight mass spectrometry , SIMION
Journal title :
Astroparticle Physics
Record number :
2022787
Link To Document :