Author/Authors :
Marnieros، نويسنده , , S. and Juillard، نويسنده , , A. and Bergé، نويسنده , , L. and Collin، نويسنده , , S. and Dumoulin، نويسنده , , L.، نويسنده ,
Abstract :
We present a study of a 100 g low temperature Ge detector, allowing identification of surface events down to the energy threshold. The bolometer is fitted with segmented electrodes and two NbSi Anderson insulator thermometric layers. Analysis of the athermal signals amplitudes allows us to identify and reject all events occurring in the first millimeter under the electrodes.
Keywords :
Dark matter , Surface events , bolometers , WIMP , Anderson insulators