Proof-of-principle study of a small animal PET/field-cycled MRI combined system using conventional PMT technology
Author/Authors :
Peng، نويسنده , , Hao and Handler، نويسنده , , William B. and Scholl، نويسنده , , Timothy J. and Simpson، نويسنده , , P.J. and Chronik، نويسنده , , Blaine A.، نويسنده ,
Abstract :
There are currently several approaches to the development of combined PET/MRI systems, all of which need to address adverse interactions between the two systems. Of particular relevance to the majority of proposed PET/MRI systems is the effect that static and dynamic magnetic fields have on the performance of PET detection systems based on photomultiplier tubes (PMTs). In the work reported in this paper, performance of two conventional PMTs has been systematically investigated and characterized as a function of magnetic field exposure conditions. Detector gain, energy resolution, time resolution, and efficiency were measured for static field exposures between 0 and 6.3 mT. Additionally, the short-term recovery and long-term stability of gain and energy resolution were measured in the presence of repeatedly applied dynamic magnetic fields changing at 4 T/s. It was found that the detectors recovered normal operation within several milliseconds following the end of large pulsed magnetic fields. In addition, the repeated applications of large pulsed magnetic fields did not significantly affect detector stability. Based on these results, we implemented a proof-of-principle PET/field-cycled MRI (FCMRI) system for small animal imaging using commercial PMT-based PET detectors. The first PET images acquired within the PET/FCMRI system are presented. The image quality, in terms of spatial resolution, was compared between standalone PET and the PET/FCMRI system. Finally, the relevance of these results to various aspects of PET/MRI system design is discussed.
Keywords :
PMT , Field-cycled MRI , PET , Dual modality , Small animal imaging
Journal title :
Astroparticle Physics